小男孩‘自慰网亚洲一区二区,亚洲一级在线播放毛片,亚洲中文字幕av每天更新,黄aⅴ永久免费无码,91成人午夜在线精品,色网站免费在线观看,亚洲欧洲wwwww在线观看

分享

An ideal blood mimicking fluid for doppler ultrasound phantoms.

 yezongying 2013-09-07

For evaluating the performance of Doppler equipment, blood itself might appear to be the best fluid to use. However, there are a number of difficulties associated with using blood and its components. It is a potential biohazard and care must be taken to minimize this risk. The shelf life of blood is limited and erythrocytes in vitro are easily damaged. This prevents the use of blood as a standard fluid in long term studies of measurements and quality control. The rheological and acoustical properties of blood are likely to be different at room temperature than at 37?C and it is an additional complication to run flow models at the higher temperature.

Human blood is non-newtonian, which means that the viscosity is strongly dependent on shear rate. The dependence on shear rate is thought to occur in small vessels. In large vessels, it is acceptable to consider that blood is newtonian.

A blood-mimicking fluid (BMF) used in Doppler ultrasound (invitro) should ideally have similar properties to real blood. The ultrasound scanner would then receive an equivalent echo signal to that from blood.

Details of relevant acoustical properties of BMF must be considered carefully. Speed of sound, backscattered power, attenuation and also density, viscosity and particle size are important factors which should be examined in primary stage of manufacturing a BMF. Having a correct speed of sound is necessary in applications such as volume flow measurements where the vessel cross sectional area is being measured. The backscatter power is affected by shear rate, haematocrit and by turbulence.[1] Attenuation is also one of the parameters which will affect the relative weighting of Doppler signals from various parts of the vessel lumen. In real blood the attenuation is small (0.9 dB cm?1 at 3.5 MHz)[1] and it should not therefore be significant in any blood analogue.

The important rheological parameters are the density, viscosity and the particle concentration or haematocrite.

This paper describes a BMF in which the scatters are nylon particles of similar dimensions to erythrocytes and which closely models the properties of real blood.


Materials and Methods
Preparation of BMF

Three different kinds of BMF were selected for experimental study; one of them (sBMF) which seems to be more reliable in mimicking blood was studied in more details.

In sBMF ultrafine polyamide particles (Orgasol ELF Atochem, Paris) of diameter 5 mm, were used as the scattering particles. The Orgasol particles had a specified nominal density of 1.03 g cm?3. The sBMF was prepared by mixing the required amount of Orgasol (1.82%) with a fluid base of (% weight): pure water (83.86%); pure glycerol (10.06%); Sigma D4876 dextran of average molecular weight 185000 Dalton (3.36%); ICI synperonic N surfactant(0.9%). The sBMF was filtered through a 30 mm sieve to remove any residual clumps. The fluid base water/glycerol proportion principally determined the speed of sound of the sBMF and the fluid density. The density of sBMF was matched to the nylon particle density so that the particle was naturally buoyant. The dextran increased the viscosity of the sBMF and had no significant effects on other physical or acoustic properties. Surfactant was used to wet the Orgasol particles. The sBMF was easy to prepared and the typical preparation time was 4-5 h.

After preparing, two important acoustic properties (velocity and attenuation) were measured. The sBMF was placed in a Plexiglas container of size 15 ? 15? 15 cm. Two 5 MHz transducers (Acuson L558) were located face to face in two opposite side of the container, one of them as a transmitter and one of them as a receiver [Figure 1]. The container was filled with the sBMF (or distilled water); before each set of measurements the sBMF (or water) was degassed by using a vacuum pump.

For evaluating the system before measuring the sound properties of sBMF, the sound velocity of water in our experimental set up device was measured.

The speed of sound (SS) can be calculated using following equation:

Cm=d/t

With Cm = SS in the material inside the water, d = thickness of material and t = time.

Velocity measurement of water was repeated several times with checking temperature of fluid at time (to determine maybe even try heating it up to see if velocity changes a lot with temperature). Also attenuation measurements have been done in water to see what the losses are from the walls of the Plexiglas container.

Then the acoustic velocity in sBMF was measured, the comparison of our experience with results of the original study shows that both results are similar and are in recommended range of IEC [Table 1].

Experimental flow model

Flow phantoms may provide steady or pulsatile flow. The purpose of the design should be considered in advance. In designing a flow phantom; the type of fluid, tubing and pump are very important and should be selected to have realistic velocity and attenuation values.

A preliminary steady flow model has been built and calibrated, using both constant pressure with no flow and constant pressure with flow with a range of tube sizes.

Many of the different particles which have been used in different blood mimicking fluids have density more than the red blood cells (RBC) which is undesirable. The fluid should have the same haemodynamic properties as well as physical properties of human blood. It is likely that the most critical for BMF is backscatter and its value for Orgasol is very closely to the IEC standard [Table 2].

The attenuation and velocity of selected BMF was measured in our lab for setting up the same condition of the original study and preventing of any systematic error. Figure 1 shows the block diagram of experimental set up for measurement of attenuation and velocity in selected BMF and water.

Among the papers that introduced the different kinds of BMFs, the BMF which is introduced[2] has been selected (sBMF) and prepared based on their recipe and some modification has been done as following:

By preparing dextran to Sigma D5501, sBMF2 with a higher viscosity was prepared.

sBMF3 with an antifungal agent (NaN3) was also prepared.

Tube

An ideal tube material should have acoustic impedance close to fluid and vessels wall as well as good elasticity. Also it should be easily formed and available in various in diameters and with uniform wall thickness. In phantom the acoustic properties should be similar to BMF with the same physical properties with no extra sound reflection. Most of the systems designed for checking the accuracy of flow measurement have been made to facilitate calibration checks. Therefore relatively large vessels and flow rates are used throughout. In this study much smaller vessels and flows will be needed. The acoustic properties of different tube materials have been summarized in Table 3.

Pump

Two different type of flow phantom have been described in the literature:[3]

  • constant flow phantom
  • pulsatile flow phantom

A pulsatile flow system, provides a more realistic model of arterial flow but it makes the interpretation of the resulting Doppler signal difficult,[4] because velocities and velocity profiles may not be known. A constant flow phantom provides a most reliable but less realistic option.

It is important to isolate the pump from the flow model in order to minimise vibration of the tubing. For producing stable flow in smallest vessels a small pump with large length of vessels is needed (i.e.: pump delivering ~0.400 ml per revolution).[5]

The pump which has been used is (EHEIM Nr. 381. Germany) and the flow velocity was controlled by varying the speed of the pump with a potentiometer, so the pump delivery was collaborated for each pump setting.


Results and Discussion

The calculation of sound velocity in distilled water and sound velocity and attenuation in the selected BMF is given in Table 1.[2, 6]

The haemodynamic and physical properties of human blood, sBMF and IEC specification have been summarized in Table 2

The acoustic properties of three different tube materials in the market at the time of the study in compare with human blood (plasma) have been summarized in Table 3.

Comparing three different materials in Table 3, it seems that by considering all three categories, acoustic properties of polyethylene tube is similar to human blood and this is consistent with most published data.

After calibration of our flow model the flow rate in four tubes with different diameters was studied. The dependence of varying the pressure using a water column on the measured pressure and flow rate for a number of diameters is summarised in Figures 2 and 3.

Since both sBMF2 and sBMF3 were prepared with almost the original amount of glycerol (10-12%) there was no change in both densities. The other properties of BMF1 were not adjusted. The compositions of all three sBMFs and physical properties are given in Tables 4 and 5.

It can be seen from the measurements in different setting the BMF3 which has been designed in our lab achieves a close match to the necessary properties of blood for in-vitro work. However the ideal BMF has not yet been produced, differences in parameter values and physical behaviour still exist between the BMF and real blood in in-vivo studies.


Conclusion

A flow test object has been developed with various options for the BMF. This test object provides flow situation that can be used to assess the capabilities of Doppler flow system. The physical properties of the studied BMF is very similar to IES standard. The BMF is easy to prepared and with adding antifungal can be used in acceptable duration of time (about 150 days) with almost no effect on physical and acoustical properties. In conclusion after modification, the BMF proved satisfactory for experimental work on Doppler ultrasound and suitable for the assessment of Doppler performance in a ultrasound system in agreement of the international standards.


Notes

Source of Support: Nil

Conflict of Interest: None declared.

References
1. Oates CP. Towards an ideal blood analogue for Doppler ultrasound phantomsPhys Med BiolYear: 1991361433421754614
2. Ramnarine KV,Nassiri DK,Hoskins PR,Lubbers J. Validation of a new blood-mimiking fluid for use in Doppler flow test objectsUltrasound Med BiolYear: 19982445199587999
3. Law YF,Cobbold RS,Johnston KW,Bascom PA. Computer-controlled pulsatile pump system for physiological flow simulationMed Bioll Eng ComputYear: 1987255905
4. Hoskins PR,Anderson T,McDicken WN. A computer controlled flow phantom for generation of physiological Doppler waveformPhys Med BiolYear: 1989341709172479955
5. Huber S,Delorme S,Knopp MV,Junkermann H,Zuna I,Van Fournier D,et al. Breast tumours: Computer assisted quantitative assessment with colour Doppler ultrasoundRadiologyYear: 19941927978018058950
6. Lubbers J,Graaff R. A simple and accurate formula for the sound velocity in waterUltrasound Med BiolYear: 199824106589809641

Figures

[Figure ID: F0001]
Figure 1 

Block diagram of experimental set up for measurement in our lab.



[Figure ID: F0002]
Figure 2 

Measurement of fluid flow in different tube diameters and different height of water column.



[Figure ID: F0003]
Figure 3 

Measurement of voltage in pressure meter in different height of water column and various tube diameters.



Tables
[TableWrap ID: T0001] Table 1 

Comparison of sound velocity in m s?1 and attenuation in dB cm?1 MHz, in water and selected BMF and references BMF


Ref. water 22?C Ref. water 37?C Water* 22?C Water* 37?C Ref. BMF 22?C BMF* 22?C
Average speed of sound(m/s) 1523.5 ? 0.5 1489.2 ? 0.5 1537 ? 5 1498.75 ? 5 1548 ? 5 1546 ? 5
Attenuation ** ** ** ** 0.05 0.06

*The measurements of this study

**Too small to be measured, BMF - Blood mimicking fluid


[TableWrap ID: T0002] Table 2 

Physical and acoustic properties of selected blood mimicking fluid compared to whole human blood.[2]


Properties IEC 1685 Draft specification Human blood 37?C Selected BMF 22?C
Scatterer RBC Orgasol ?(nylon)
Scatterer size (mm) 1050 ? 40 7 5
Hematocrit (% volume) 4 ? 0.4 45 <5
Density (kg m?3) 1570 ? 30 1053 1037 ? 2
Viscosity (mPa s) < 0.1 3 4.1 ? 0.1
Velocity (ms?1) Comparable to RBC 1583 1548 ? 5
Attenuation(dB/cm MHz) Newtonian 0.15 0.05 ? 0.01
Backscatter (dB) 0 Comparable to RBC
Fluid properties non Newtonian Newtonian

BMF - Blood mimicking fluid, RBC - Red blood cells


[TableWrap ID: T0003] Table 3 

Acoustic properties of three different tube materials in comparison with plasma.[3]


Material Density kgm?3 Velocity ms?1 Impedance* Kg (s.m2)?1
Polyethylene 980 540 1.76 ? 106
Acrylic 1190 1330 3.27 ? 106
Glass (Pyrex) 2240 3280 1.26 ? 107
Plasma 1026 1546 ~ water*

*Impedance of water at 25 ?C = 1.49 ? 106 kg/s.m2


[TableWrap ID: T0004] Table 4 

Composition (components in mass %) of three BMFs


Components sBMF sBMF2 sBMF3
Water 83.86 84.02 82.04
Glycerol 10.06 10.8 12.06
Dextran 180 kD 3.36 3.36 --
Dextran 5-40 MD -- -- 1.15
Orgasol 5 mm 1.82 1.82 1.82
Synperonic N surfactant 0.9 0.9 0.90
NaN3 -- 0.2 --

BMF - Blood mimicking fluid


[TableWrap ID: T0005] Table 5 

Physical properties of three sBMFs in 22?C


Physical properties sBMF sBMF2 sBMF3
Density (kg m?3) 1037 ? 2 1038 ? 2 1039 ? 2
Dynamic viscosity (mpa s) 4.1 ? 0.1 3.83 ? 0.1 4.7 ? 0.1
Sound velocity (m s?1) 1548 ? 5 1551 ? 5 1563 ? 5
Attenuation dB (cm. MHz)?1 0.05 ? 0.01 0.05 ? 0.01 0.06 ? 0.01
Back scattering (dB) ?36.0 ?35.3 ?36.7

BMF - Blood mimicking fluid



Article Categories:
  • Original Article

Keywords: Blood mimicking fluid, doppler ultrasound, flow test object.

    本站是提供個(gè)人知識(shí)管理的網(wǎng)絡(luò)存儲(chǔ)空間,所有內(nèi)容均由用戶發(fā)布,不代表本站觀點(diǎn)。請(qǐng)注意甄別內(nèi)容中的聯(lián)系方式、誘導(dǎo)購(gòu)買等信息,謹(jǐn)防詐騙。如發(fā)現(xiàn)有害或侵權(quán)內(nèi)容,請(qǐng)點(diǎn)擊一鍵舉報(bào)。
    轉(zhuǎn)藏 分享 獻(xiàn)花(0

    0條評(píng)論

    發(fā)表

    請(qǐng)遵守用戶 評(píng)論公約

    類似文章 更多